PHOTOMETER SPECIFICATION
◎
Designed for the purpose of routine clinical bio-chemistry
tests, coagulation time tests.
◎
Hermetic solid-state optical system, low
contamination of dust and dampness, long life time of
hard membrane filters than soft ones.
◎
Light splitting after cuvette, separate
signal amplifier for each wavelength, filters select
completed electronically, no mechanical movements, fast
measurement and precise results, especially useful for
bichromatic measurements.
◎
Dedicated membrane keyboard, menu
selective with operating guiding. Check on the instrument
status at any time with self-test function.
◎
Back-light 240-128 matrix LCD, character,
ASC code and curve simultaneously displayed, readable even
if in the dim light. Displaying in English.
◎
Thermal printer, 58mm, fast speed and low
noise, optional external printer.
◎Test
methodology: end-point (absorbance, linear concentration,
nonlinear concentration, with bichromatic test), fixed time,
kinetic, coagulation time test.
◎
Light source: 12V 20W, Tungsten halogen
lamp, life time
³2000
hr.
◎
Optical system: 6 hard membrane filters,
340 405 505 546 578 630nm, wavelength precision
±2nm.
◎
Abs. range: 0—2.5000Abs, resolution:
0.0001Abs, drift<0.003Abs/hr after 10 minutes warm up.
◎
Real time clock displayed.
◎
Precision: Abs=2.0000 CV<1%, Abs=0.5000
CV<0.5%
◎
Temperature: PELITIER control,
25,30,32,37
±0.1°C.
◎
Cuvette: Mico-flow cell 25mL.
◎
Power: 220V 50HZ/110V 60HZ,
£80VA.
◎
Standard Rs232 serial port.
◎
Dimension: 420(L) 320(W) 170(H)mm
◎
Weight: 11kg(gross),10kg(net)