Automatic Urine Analyser
User Guide of Automatic Urine Analyzer

Contents
Ⅰ.
General Introduction
2
Ⅱ.
Instrument Structure
2
Ⅲ.
Instrument Functions & Technical Parameters
3
(ⅰ)
Functions
3
(ⅱ)
Technical Parameters
3
Ⅳ.
Operating Guide
4
(ⅰ)
Installation
4
1. Installing the Test paper holder
4
2. Installing the Printing paper
5
3. Charging the Battery
6
(ⅱ)
Operating Instructions
6
1. Urine Test
7
2. Continuous Urine Tests
10
3. Pregnancy Test
11
(ⅲ)
Settings
13
1. Setting the Date
13
2. Setting the Time
13
3. Setting the Testing Items
13
4. Print Settings
14
(ⅳ)
Data Printing
14
(ⅴ)
Data Queries
15
(ⅵ)
Erasing
15
Ⅴ.
Instrument Maintenance
16
Ⅵ.
Notice
16
Ⅶ.
Problems & Troubleshooting
16
Ⅷ.
Waste Disposal
17
Ⅸ.
Package List
17
User Guide of
Automatic Urine Analyzer
Ⅰ.
General Introduction
The urine
analyzer is a popular routine urine analyzer, which can
analyze white cells, tumor cells, nitrite, proteins, pH,
proportion, ketone, bilirubin, urobilinogen, glucose,
vitamin C, creatinine, calcium, etc. in urine automatically.
Either 10 or 11 of the above items can be chosen to analyze
according to needs. There are two types of result-output
ways: liquid crystal display output and thermosensitive
printing output. 60 test data can be saved to the memory.
Besides, this instrument has the function of pregnancy test
with our quick pregnancy test paper. Both power supply ways:
direct current charging and alternating current changing can
be used, so that it won’t stop working during the power cuts
or in the ambulance. In addition, it can be used in both
individual sample tests and continuous batches of samples
tests. It is very simple and convenient to operate and use
this instrument with the operating guide shown on the
display panel. This urine analyzer is suitable for use in
all hospitals, local surgeries, ambulance, school doctor
rooms, etc.
Ⅱ.
Instrument Structure
This section describes the structure and
names of each component of the urine analyzer C-L 2.

1. Buttons 2.
Liquid Crystal Display Panel 3. Paper Cover 4. Charging
Indicator 5. USB Connection Indicator 6. Test Paper Holder
7. Photosource Adjustment Block 8.Test Paper Holder Access
9. Test Paper Trough
10. Test Paper
Locating Graticule 11. Power Supply Adapter 12. Paper
Tear-off.
Diagram 1. Vertical View

13. Power Switch 14.USB Socket 15. Power
Supply Adapter Socket 16. Top Cover 17. Lower Cover
18. Battery Cover 19. Screws
Diagram 2. Back
View and Bottom View
Ⅲ.
Instrument Functions &Technical Parameters
(ⅰ)
Functions:
1.
To test: white cells, tumor cells, nitrite, proteins,
pH, proportion, ketone, bilirubin, u robilinogen, glucose,
vitamin C, creatinine, calcium, etc. in urine. Apart from
the above items, HCG in urine can be examined for pregnancy
test.
2.
Both printout and panel display are available.
3.
The instrument is able to work continuously.
4.
A power supply adapter and a battery pack are
equipped for either direct or alternating mains supply. It
can continue working for 5 hours and continuously testing
120 samples during a power failure.
5.
60 test data can be saved to the memory and inquired,
or printed out on occasion.
(ⅱ)
Technical Parameters
1.
Power Supply: DC9-10V (A 9V or 10V alternating
current supply adapter is equipped with the instrument)
2.
Operating Conditions: Humidity: <80%
Temperature: 25℃±4℃
3.
Speed: >55 times/hour
4.
Dimensions: 265mm×217mm×96mm(Length×Width×Height)
5.
Weight: ≈1.8kg
6.
Life-span: The mainframe: Approx. 5-10 years
The battery pack:
Approx. 2-3 years (OZHainable in the manufacturer)
7.
Standard Parameters of the 13-Item Tests
|
Items |
Contraction |
Output
Data |
|
White
Cells |
LEU |
- 15 70 125 500
Leu/μL |
|
Tumor
Cells |
BLD
|
- 10 25 80 200
Ery/μL |
|
Nitrite |
NIT |
- + |
|
Urobilinogen |
URO |
3.2、16、33、66、150、200
μmol/L |
|
Proteins |
PRO |
- ±
+1 +2
+3 +4
0.15 0.3
1.0 3.0 10.0
g/L |
pH
|
|
5 6 7 8 9 |
|
Proportion |
S.G
|
1.000 1.005 1.010 1.015 1.020 1.025 1.030 |
|
Ketone |
KET |
- ± +1 +2 +3 +4
0.5 1.5 4 8 15
mmol/L |
|
Bilirubin |
BIL |
- +1 +2 +3
8.5
35 100 μmol/L |
|
Glucose |
GLU |
- ± +1 +2 +3 +4
2.8 5.5 14 28 55
mmol/L |
|
vitamin C |
Vc |
0 0.6 1.4
2.8 5.6
mmol/L |
|
Creatinine |
CRE |
0 0.9 1.8
4.4 22 >22
mmol/L |
|
Calcium |
Ca |
0 0.5 1.25 2.5 12.5 25
mmol/L |
Ⅳ.
Operating Guide
(ⅰ)
Installation
1.
Installing the Test Paper Holder
Insert 2/3 of the
test paper holder into the instrument through the access.

Diagram 3
2.
Installing the Printing Paper
(1). Open the
paper cover following the method of Diagram 4 and 5

Diagram 4
Diagram 5
(2). Prepare some printing paper. Follow
Diagram 6 and 7, push the pale green button with your right
hand. Then insert the printing paper into the printer and
pull out the end of paper approximately 5 cm as show in
diagram.9

Diagram 6
Diagram
7

Diagram 8
Diagram 9
(3) Be sure that
the end of paper stretched out through the paper access, and
close the paper cover as shown below.

Diagram 10
Diagram 11
3.
Charging the Battery
(1)
Plug in the instrument with the power supply adapter,
and insert the input plug of the
adapter into the 220V 50HZ alternating electrical
outlet.
(2)
The charging indicator will flash while the battery
pack is charging. It takes approximately 13 hours to fully
charge the battery pack for the first time. The second
charging should only last till the indicator goes out. As
soon as the battery is fully charged, the instrument can be
put to use.
(ⅱ)
Operating Instructions
Now you have
prepared the instrument, you are ready to operate it. Turn
it on approximately 5 minutes before use every time.
This instrument
will self check automatically when it is turned on. The main
menu will then be displayed on the display panel.

Diagram 12
The date and time
has been set during production. Reset it by pressing
[Settings] button if you find it is incorrect.
Allow 5 minutes
for the preheating before the urine test started.
1.
Urine Test
Before the test
started, be sure that the cursor is at [Urine Test] position
([Urine Test] is the Factory default). And correct the
number of testing items (13 items is the factory default).
Start the test when the main menu is confirmed.
As soon as the
urine sample and the test paper is prepared respectively,
follow the below procedures for the test: hold the prepared
test paper with your right hand, and press the Yes button
with your left forefinger so that [Ready for the Urine Test]
will be displayed as you do so far. Immerse the test paper
in the urine sample as soon as [Start Urine Immersion] is
displayed. And take it out when you see [Urine Immersion
Finished] on the panel.




Diagram 13
It takes
approximately 2 seconds for the immersion. Blot excess urine
on the side of the test paper with a paper towel after it is
taken out.

Diagram 14
Put the test paper in the trough of the
test paper holder. When it counts down to the 10th
second on the timer displayed on the panel as shown below.

Diagram 15
Ensure that the test paper clings to the
surface of the trough. If taking the 13-item test, line up
the outside line of the second test paper, which is near the
test paper end, with the locating indicating line on the
test paper holder.

Diagram 16
When taking the
10 or 11-item test, you should get the black line on the
test paper end and the locating indicating line on the test
paper holder to line up. Vide Diagram 17.

Diagram 17
After putting the
test paper into the holder, it will be checked when it
counts down to the 8th second. As [Test Paper
has been Confirmed] is displayed, the test paper will be
sent into the instrument with the test paper holder
automatically. The holder will move outwards when it shows
[Test Started] on the panel. i.e. the instrument begins to
scan and examine the test paper from outside to inside. The
test report will be printed out soon after [Test Finished]
is displayed.



Diagram 18

Diagram 19
2. Continuous
Urine Tests
If taking the
continuous tests of several urine samples, you should
prepare a test paper for one urine sample as soon as the
previous test paper is confirmed. When [Ready for the Urine
Test] is displayed, you should start the test for the next
sample by following the procedures of Urine Test. Carry on
the continuous tests by repeating this process. If you need
stop it, do not insert a new test paper when the previous
test finishes. Then the main menu will be shown on the panel
automatically. Press [Yes] button to start the next test, if
more tests are needed.

Diagram 20
3. Pregnancy
Test
This instrument
can be used for pregnancy test by examining urine samples.
This test must be carried out with the 4mm-5mm quick
pregnancy test paper produced by Shanghai Batong Institute
of Biotechnology. If using the test paper from other
manufacturers, errors in the test results will be caused by
mismatches in the structure and dimensions.
Set the
instrument every time before taking pregnancy tests. Press
the ▽
button to move the cursor to [Pregnancy Test]. Vide Diagram
21.

Diagram 21
Hold the quick
pregnancy test paper with your right hand while press [Yes]
button with your left hand. There will be [Ready for the
Pregnancy Test] displayed on the panel. When it turns to
[Start Urine Immersion], immerse the test paper in the urine
sample (The liquid level of urine sample can not exceed the
MAX line). Take the test paper out as [Urine Immersion
Finished]. Vide Diagram 22.
Put the test
paper in the trough of the test paper holder with its right
side up. And push it to the innermost of the trough when it
says [Insert the Test Paper] on the panel.



Diagram 22

Diagram 23
Take the test paper out of the holder,
and press the Yes button. When the test finished, press
[Yes]. As [Test Paper is Taken out] shown on the panel, the
pregnancy test report will soon be printed out.
![文本框: .Test Finished
Press [Yes] when the Test Paper is Taken out](Urine%20Analyser.files/image015.gif)

Diagram 24
(ⅲ).
Settings
Settings are
mainly used for time/date correction, testing-item selection
and print settings.
1.
Setting the Date:
Use the ▽
button to select [Settings]. Press the Yes button twice, so
that the field [Year] is selected with a horizontal cursor
above.

Year
Month Day Hour Minute specification Print Settings
Number
Diagram 25
Use the ▽
or △
button to select the correct date. Confirm with [Yes]
button. Then the cursor will move to [Month]. Press [Yes]
button again to turn the cursor to [00]. Set the right mouth
with the ▽
or △
button and press the Yes button to confirm. Ditto to the
day’s setting. To return to the main menu, press the No
button.
2. Setting the
Time:
Move the cursor to ‘Settings’ by pressing the button ▽.
And press the Yes button 8 times to move the horizontal
cursor to [Hour]. See Diagram 26.

Diagram 26
Adjust the value
using the button ▽
or △,
and press the Yes button to confirm. The same to the minute
setting. Return to the main menu by pressing the button
[No].
3. Setting the
Testing Items:
The factory default for the number of testing items is 13.
To use for the 10 or 11-item test, you need to reset before
test starts.
Select
[Settings], and press the Yes button 12 times to move the
horizontal cursor to [13T]. Use the ▽
or △
button to select [10T] or [11T] needed. Confirm with the Yes
button likewise.

Diagram 27
4. Print
Settings:
[Printing Synchronization] (P:ON) is the factory default for
the printing mode. i.e. the test report will be printed out
automatically after each test. If the printout is not
needed, follow the same procedures to move the horizontal
cursor to [P:ON], and select [P:OFF] with the ▽
button, and use the button [Yes] to confirm.

Diagram 28
(ⅳ).
Data Printing:
The test report
will be printed out automatically soon after each test since
[P:ON] is the factory default printing mode. Therefore, the
Data Printing will only be used if the printing mode was [P:OFF]
or reprinting is needed.
Use the ▽
button to select [Data Printing]. Press [Yes] to print out
the test report with its number displayed on the panel as
show below.

Diagram 29
(ⅴ).
Data Queries:
In the continuous
tests, all of the test results can be printed out, while
only the last result will be shown on the display panel. To
query the rest results of other numbers, use the data
querying function.
Select [Data
Queries], press the Yes button to move the horizontal cursor
to [NO:000]. Adjust the right number with the ▽
or △
button. If you need to print out the result, select [Data
Printing] again.



Diagram 30
(ⅵ)
Erasing:
Due to that 60
test data can be saved, and the 60 numbers file in serial
order (e.g. If 25 samples are tested on the first day, the
serial numbers will start from the 26th on the
next day), erase the record before shutting the instrument
down every day to keep the serial number starting from the 1st.
Use the ▽
button to select [Erasing], all the test data will be
deleted.
Ⅴ.
Instrument Maintenance
Follow the procedures below to maintain
the instrument.
1. Remove the power supply adapter during
the usage of the instrument when the battery pack is fully
charged. When the battery is low, recharge the battery with
the adapter till the charging indicator is out. Be sure to
charge it fully before shutting the instrument down.
2. Take out and
clean the test paper holder after each use. Press [Yes] in
the waiting state. Power off when the holder moves out. Draw
out the holder, and rinse the test paper trough under the
tap with a small brush to get rid of dirt. Mop up water with
a paper towel.
3. Gently wipe
dirt and dust off the instrument to keep it clean.
Ⅵ.Important
Please notice the
following items when use.
1. Lay the
instrument in a dry place. Moisture and corrosive will cause
damage to it.
2. Avoid direct
sunlight.
3. Ensure that
the operating temperature is 25℃±4℃.
Otherwise, errors will be caused.
4. Please keep
strictly to the operating guide (e.g. urine immersion time,
test paper locating, etc. to minimize errors.
5. When rinsing
the test paper holder, check the locating hole in the side
of the holder to avoid blockages in it (observe the locating
hole towards the light). Dredge it if blocked by water or
dirt to avoid inward dislacation, etc.
6. Please do not
turn on and off the instrument frequently. Ensure that the
interval between is more that 5 minutes to avoid systems
disorder.
7. Please do not
press two or more than two keys simultaneously to avoid
malfunctions.
Ⅶ.Problems
& Troubleshooting
Highly integrated technology is used in
this instrument. And the structural design of its driving
part is rather simple. As a result, malfunction is
relatively less when use.
There are several common problems and the
relevant troubleshooting listed below:
|
Problem |
Cause(s) |
Solution(s) |
|
Panel
display disorder |
Frequent
Operation |
Turn off
the power, allow 5 minutes before restarting the
instrument |
|
Printing
paper jam |
|
Pull the
end of recording paper out of the paper access
approximately 5 cm, and close the paper cover. |
|
Printing
paper deflected |
The roll
paper is not in the center of the paper chamber |
Open the
paper cover. Push the pale green button, and move
the roll paper to the paper chamber center. Release
the button and pull the paper end out. Close the
paper cover finally. |
|
Movement
disorder of the test paper holder |
Blockages
in the locating hole |
Turn off
the power first. Screw the mounting screws off to
open the top cover slightly. Check if the locating
hole is blocked by water or dirt as the test paper
holder is taken out. If so, dredge it with a thin
iron wire. Reinstall the instrument then. If it
still does not work, please consult the manufacturer
rather than open the instrument presumptuously. |
Others: The maintenance to other
malfunctions must be conducted by the manufacturer or the
professionals.
Ⅷ.
Waste Disposal
The used test
paper should be disposed of as nonrecoverable medical waste.
Ⅸ.
Package List
Urine Analyzer C-L 2 Mainframe 1
Power Supply Adapter 1
Test Paper
Holder 1 Battery Pack (installed in the
mainframe) 1
Roll Recording
Paper 1 13-Item Urine Test
Paper 1