Home
Catalogue
Contact
us
One Step H
Pylori Test
Read
the instruction before performing assays. Follow the
procedures during testing.
SUMMARY
The infection of
Helicobacter pylori 1 (H. pylori, spiral- shaped
bacterium) is associated with a variety of
gastro-intestinal diseases, such as stomach ulcer,
chronic active gastritis and gastric, duodenal, and gastric
denocarcinoma2,3,4,5,6,7,8. Antibodies to H. pylori are
developed in Individuals infected with H. pylori as a
serological response 9,10. Detecting these specific
antibodies to H. pylori can be used as a qualitative assay
in the diagnosis of H. pylori infection. It may be used
either as an adjunct to endoscopy or as an alternative
measure in symptomatic patients.
PRINCIPLE
The Diagnostics H.
pylori one-step assay is a lateral flow, chromatographic
immunoassay (CIA), containing filter membrane
pre-coated with H. pylori specific antigens and colored gold
colloidal reagents labeled with H. pylori specific
antigens. There are two regions on the membrane of the test
strip, test region (T) and control region (C). A T line,
the burgundy colored band in the test region of membrane,
will be developed rapidly (from 30 seconds to 5 minutes)
when antibodies to H. pylori present in specimen. If
antibodies to H. pylori are not present, no T line will be
developed in the test region. A C line, a burgundy colored
band in the control region of the test strip, should always
appear regardless of the presence of antibodies to H.
pylori, serving as a internal qualitative control of the
test system.
REAGENTS and
MATERIALS
1. Test Device:
Each contains a test strip.
2. Dilution
buffer.
STORAGE
The kit should be
stored at room temperature (15-30°C) or refrigerator
(2-8°C) in the sealed pouch with desiccant for the
duration of the shelf life (18 months).
SPECIMEN
COLLECTION
1.
Both fresh
whole blood and serum specimen may be tested with this
assay. Following standard clinical procedure to collect specimens.
2. Finger-stick
is recommended to collect fresh whole blood specimen for
this assay. Wipe fingertip with alcohol, then prick
fingertip with a lancet in a quick motion.
3. Serum
specimens can be stored at room temperature (20°C to 28°C)
for 4 hours, at 2o-8oC for 48 hours,
and at -20°C or lower for prolonged storage. Repeatedly
frozen and thawed specimens are not recommended for this
assay.
4. Any sediment
in serum specimens should be removed by centrifugation.
Avoid using any turbid specimens, which may be contaminated
by microorganisms.
PRECAUTIONS
1. FOR IN VITRO USE ONLY.
2. Do not use
test kit beyond expiration date.
3. Do not open
the test device foil pouch until you are ready to perform
the test.
4. Icteric,
lipemic, hemolysed, heat treated and contaminated sera may
cause erroneous results.
PROCEDURE
1.
Refrigerated
specimens or other materials should be equilibrated to room
temperature right before performing assay.
2. Remove the
device from its protective pouch. Label the device with
specimen identification.
3. Add 40ml
fresh whole blood (or 20ml serum) of specimens into the
sample well.
4. Add four
drops (about 200 ml) of dilution buffer into the sample
well.
5. Positive
results may be observed as soon as 30 seconds after the
specimen reaches the results window, depending on
the concentration of antibodies to H. pylori in the
tested specimen. To confirm negative results, total of 5
minutes, a complete reaction time is required.
INTERPRETATION OF RESULTS
.files/image003.gif)
Positive:
Two pink-rose bands appear: one in the test region (T) and
in the control region (C).
Negative:
One rose-pink band appears in the control region (C), with
no band in the test region (T).
Invalid: There is no distinct color band visible both in the test
region and in the control region, or there is a visible band
only in the test region and not in the control region. The
result is invalid due to deterioration of the test or
improper test procedure. Repeat the test with a new Testing
Device.
LImitation
1. This test is
a qualitative assay. To confirm diagnosis of gastritis
and/or peptic ulcers, clinical findings need to be
considered.
Serum
samples from patients infected with C. Jejuni may have a
low cross-reactivity with this test.
A negative
result can not rule out H. pylori infection.
Because the antibody to H. pylori may be absent or
may be insufficient to be detected at the time of
testing.