One Step Cassette Style AFP Test

                                  

Introduction

One step AFP cassette test is a rapid direct binding test for the visual detection of alphafetoprotein in serum as an aid in the diagnosis of primary hepatocellular carcinomas, testicular teratocarcinomas and neural tube defects. Test results are read visually without any instrument.

One step afp cassette test is based on the principle of sandwich immunoassay for determination of afp in serum. .monoclonal and polyclonal antibodies were employed to specifically identify afp with high sensitivity.  This one step test only takes about 10 minutes.the sensitivity of the test can reach to 25ng/ml.

 

Specimen collection

For serum, collect blood into container without anticoagulant. Allow the blood to clot and separate the serum from the clot. Use the serum for testing.

If the specimen cannot be tested on the day of collection, store the serum specimen in a refrigerator or freeze. Bring the specimen to normal room temperaure before testing. Do not freeze and thaw the specimen repeatedly.

 

Test procedure

 

1.review specimen collection instructions, test device, patients samples and controls should be brought to room temperature(18-30℃) prior to testing.

2.remove the test device from its protective pouch(bring the device to room temperature before opening of the pouch to avoid condensation of moisture on the membrane). Label the device with patient or control identifications.

3.draw sample to the line marked on the pipettor  (approximately 0.2ml ) . Dispense entire contents into the sample well . Alternatively, hold the pipette in a straight up and down position - not at an angle. Dispense 6 drops (approximately 0.2ml ) of the specimen or control into the sample well (diag. 1) for each sample or control , use a separate pipettor and device.

4.wait for 10-20 minutes and read results. And do not read results after 30 minutes.

 

 

Interpretation of results

 

  Negative: only one pink colored band appears on the control regioncno apparent band on the patient test region(t).

  Positive:  in addition to a pink colored control band , a distinctPink colored control band will also appear in the patient test region(t)

  Invalid :  a total absence of color in both regions is an infication of procedure error and/ or that test reagent deterioration has occurred.

 

 Ne